Resultados Refrativos e Visuais do Mundo Real da Lente Intraocular Tórica Rayone® e Desempenho da Calculadora Raytrace®

Autores

  • João Alves Ambrósio Serviço de Oftalmologia, Centro Hospitalar de Entre o Douro e Vouga, Santa Maria da Feira, Portugal https://orcid.org/0000-0003-2009-5126
  • Vítor Miranda Serviço de Oftalmologia, Centro Hospitalar de Entre o Douro e Vouga, Santa Maria da Feira, Portugal https://orcid.org/0000-0003-4480-5225
  • Catarina Aguiar Serviço de Oftalmologia, Centro Hospitalar de Entre o Douro e Vouga, Santa Maria da Feira, Portugal https://orcid.org/0000-0003-2487-4231
  • Inês Almeida Serviço de Oftalmologia, Centro Hospitalar de Entre o Douro e Vouga, Santa Maria da Feira, Portugal
  • Miguel Ruão Serviço de Oftalmologia, Centro Hospitalar de Entre o Douro e Vouga, Santa Maria da Feira, Portugal https://orcid.org/0000-0002-0858-0164
  • Lígia Figueiredo Serviço de Oftalmologia, Centro Hospitalar de Entre o Douro e Vouga, Santa Maria da Feira, Portugal
  • João Chibante Pedro Serviço de Oftalmologia, Centro Hospitalar de Entre o Douro e Vouga, Santa Maria da Feira, Portugal

DOI:

https://doi.org/10.48560/rspo.33263

Palavras-chave:

Catarata, Astigmatismo, Implantação de Lente Intraocular, Lente Intraocular

Resumo

INTRODUÇÃO: Considerando a crescente disponibilidade de lentes intraoculares (LIO) tóricas, são necessários dados de resultados refrativos do mundo real para auxiliar no processo de decisão e escolha da LIO na cirurgia de catarata. O presente trabalho avalia os resultados refrativos associados à implantação da LIO RayOne® tórica, o desempenho da calculadora Raytrace®OnlinePremium e preditores de sucesso cirúrgico.
MÉTODOS: Estudo retrospetivo, observacional, de 163 olhos, submetidos a cirurgia de catarata com implante de LIO RayOne® tórica, de maio de 2022 a março de 2023, no Centro Hospitalar de Entre o Douro e Vouga.
RESULTADOS: Foram incluídos 163 olhos (120 doentes, 67 mulheres, idade média 73,1±9,0 anos). A queratometria média foi 44,13±1,90 dioptrias (D) e o cilindro corneano -1,99 [-2,40;-1,63] D. Utilizaram-se os biómetros (1)IOLMaster® 500(Zeiss), 92 (56,4%) olhos; (2)AL-Scan®(Nidek), 47 (28,8%) olhos; (3)PacScanPlus®(SonomedEscalon) associado a queratómetro TonorefII®(Nidek), 13 (8,0%) olhos; e (4)Pentacam®(Oculus), 11 (6,7%) olhos. Considerou-se um valor médio de astigmatismo corneano posterior (ACP) em 61 (37,4%) olhos. A melhor acuidade visual corrigida, o equivalente esférico (EE), a esfera e o cilindro foram 0,30 [0,22;0,40] logMAR, -1,17±3,31 D, -0,49±3,00 D e -1,38±2,02 D no pré-operatório e 0 [0;0,10] logMAR, -0,41±0,65 D, -0,07±0,58 D e -0,68±0,78 D no pós-operatório, respetivamente. A diferença entre a refração manifesta e a previsão da calculadora foi EE -0,19±0,63 D, esfera 0,08±0,58 D e cilindro -0,55±0,78 D. Não se observaram diferenças estatisticamente significativas na refração manifesta ou entre esta e a previsão da calculadora nos grupos com e sem inclusão do ACP. Verificaram-se diferenças no erro de previsão da calculadora entre biómetros – EE: 2 e 3 (-0,33±0,67 D vs 0,29±0,69 D, p=0,01), 3 e 4 (0,29±0,69 D vs -0,45±0,91 D, p=0,02); cilindro 3 e 4 (-0,44±0,91 D vs -1,29±0,65 D, p=0,04), 1 e 4 (-0,61±0,76 D vs -1,29±0,65 D, p=0,03). Houve rotação superior a 10º em 5 casos, reposicionamento num deles, rotura da cápsula posterior num caso, e erro humano no cálculo ou encomenda da LIO em dois casos.
CONCLUSÃO: A biometria ultrassónica tem maior tendência de hipermetropização e o Pentacam® de subcorreção do cilindro. No mundo real, a LIO RayOne® tórica e a calculadora Raytrace®OnlinePremium demonstram bons resultados refrativos e visuais, segurança, precisão na refração estimada e estabilidade rotacional.

Downloads

Não há dados estatísticos.

Referências

Brogan K, Diaper CJM, Rotchford AP. Cataract surgery refractive outcomes: representative standards in a National Health Service setting. Br J Ophthalmol. 2019;103:539-43. doi: 10.1136/bjophthalmol-2018-312209.

Lundström M, Dickman M, Henry Y, Manning S, Rosen P, Tassignon MJ, et al. Risk factors for refractive error after cataract surgery: Analysis of 282 811 cataract extractions reported to the European Registry of Quality Outcomes for cataract and refractive surgery. J Cataract Refract Surg. 2018;44:447–52. doi: 10.1016/j.jcrs.2018.01.031.

Kessel L, Andresen J, Tendal B, Erngaard D, Flesner P, Hjortdal J. Toric intraocular lenses in the correction of astigmatism during cataract surgery: a systematic review and meta-analysis. Ophthalmology. 2016;123:275-86. doi: 10.1016/j.ophtha.2015.10.002.

Khan MI, Muhtaseb M. Prevalence of corneal astigmatism in patients having routine cataract surgery at a teaching hospital in the United Kingdom. J Cataract Refract Surg. 2011;37:1751–5. doi: 10.1016/j.jcrs.2011.04.026.

Guan Z, Yuan F, Yuan YZ, Niu WR. Analysis of corneal astigmatism in cataract surgery candidates at a teaching hospital in Shanghai, China. J Cataract Refract Surg. 2012;38:1970–7. doi: 10.1016/j.jcrs.2012.07.025.

Patel CK, Ormonde S, Rosen PH, Bron AJ. Postoperative intraocular lens rotation. Ophthalmology. 1999;106:2190–6.

Arba-Mosquera S, Merayo-Lloves J, de Ortueta D. Clinical effects of pure cyclotorsional errors during refractive surgery. Invest Ophthalmol Vis Sci. 2008;49:4828-36. doi: 10.1167/iovs.08-1766.

Rubenstein JB, Raciti M. Approaches to corneal astigmatism in cataract surgery. Curr Opin Ophthalmol. 2013;24:30–4. doi: 10.1097/ICU.0b013e32835ac853.

Mozayan E, Lee JK. Update on astigmatism management. Curr Opin Ophthalmol. 2014;25:286–90. doi: 10.1097/ICU.0000000000000068.

Visser N, Bauer NJC, Nuijts RM. Toric intraocular lenses: Historical overview, patient selection, IOL calculation, surgical techniques, clinical outcomes, and complications. J Cataract Refract Surg. 2013;39:624–37. doi: 10.1016/j.jcrs.2013.02.020.

Laurendeau C, Lafuma A, Berdeaux G. Modelling lifetime cost consequences of toric compared with standard IOLs in cataract surgery of astigmatic patients in four European countries. J Med Econ. 2009;12:230–7. doi: 10.3111/13696990903257439.

Preussner PR, Hoffmann P, Wahl J. Impact of posterior corneal surface on toric intraocular lens (IOL) calculation. Curr Eye Res. 2015;40:809–14. doi: 10.3109/02713683.2014.959708.

Reitblat O, Levy A, Kleinmann G, Abulafia A, Assia EI. Effect of posterior corneal astigmatism on power calculation and alignment of toric intraocular lenses: Comparison of methodologies. J Cataract Refract Surg. 2016;42:217–25. doi: 10.1016/j.jcrs.2015.11.036.

Novis C. Astigmatism and toric intraocular lenses. Curr Opin Ophthalmol. 2000;11:47–50. doi: 10.1097/00055735-200002000-00007.

Felipe A, Artigas JM, Díez-Ajenjo A, García-Domene C, Alcocer P. Residual astigmatism produced by toric intraocular lens rotation. J Cataract Refract Surg. 2011;37:1895–901. doi: 10.1016/j.jcrs.2011.04.036.

Jin H, Limberger IJ, Ehmer A, Guo H, Auffarth GU. Impact of axis misalignment of toric intraocular lenses on refractive outcomes after cataract surgery. J Cataract Refract Surg. 2010;36:2061–72. doi: 10.1016/j.jcrs.2010.06.066.

Lee BW, Galor A, Feuer WJ, Pouyeh B, Pelletier JS, Vaddavalli PK, et al. Agreement between Pentacam and IOLMaster in patients undergoing toric IOL implantation. J Refract Surg. 2013;29:114-20. doi: 10.3928/1081597X-20130117-06.

Park DY, Lim DH, Hwang S, Hyun J, Chung TY. Comparison of astigmatism prediction error taken with the Pentacam measurements, Baylor nomogram, and Barrett formula for toric intraocular lens implantation. BMC Ophthalmol. 2017;17:156. doi: 10.1186/s12886-017-0550-z.

Wade M, Steinert RF, Garg S, Farid M, Gaster R. Results of toric intraocular lenses for post-penetrating keratoplasty astigmatism. Ophthalmology. 2014;121:771–7. doi: 10.1016/j.ophtha.2013.10.011.

Stewart CM, McAlister JC. Comparison of grafted and non-grafted patients with corneal astigmatism undergoing cataract extraction with a toric intraocular lens implant. Clin Exp Ophthalmol. 2010;38:747–57. doi: 10.1111/j.1442-9071.2010.02336.x.

Kersey JP, O’Donnell A, Illingworth CD. Cataract surgery with toric intraocular lenses can optimize uncorrected post-operative visual acuity in patients with marked corneal astigmatism. Cornea. 2007;26:133–5.

Cha D, Kang SY, Kim SH, Song JS, Kim HM. New axis-marking method for a toric intraocular lens: mapping method. J Refract Surg. 2011;27:375-9. doi: 10.3928/1081597X-20101005-01.

Downloads

Publicado

2024-12-26

Como Citar

Alves Ambrósio, J., Miranda, V., Aguiar, C., Almeida, I., Ruão, M., Figueiredo, L., & Chibante Pedro, J. (2024). Resultados Refrativos e Visuais do Mundo Real da Lente Intraocular Tórica Rayone® e Desempenho da Calculadora Raytrace®. Revista Sociedade Portuguesa De Oftalmologia, 48(4), 255–262. https://doi.org/10.48560/rspo.33263