Anterior Surface Measurements Correlation and Agreement Between Rotational Scheimpflug and Placido Disc-based Devices in Primary Eyes

Authors

  • F. Faria-Correia Serviço de Oftalmologia, CUF Porto, Porto, Portugal; Oftalconde, Porto, Portugal; Serviço de Oftalmologia, Hospital de Braga, Braga, Portugal; School of Medicine, University of Minho, Braga, Portugal; Rio de Janeiro Corneal Tomography and Biomechanics Study Group, Brazil
  • Sara Ribeiro Hospital de Braga, Braga, Portugal; Hospital Lusíadas, Porto, Portugal
  • Bernardo T. Lopes University of Liverpool, Liverpool, UK
  • Marcella Q. Salomão Universidade Federal de São Paulo (UNIFESP), São Paulo, Brazil
  • Renato Ambrósio Jr. Universidade Federal do Estado do Rio de Janeiro (UNIRIO), Rio de Janeiro, Brazil

DOI:

https://doi.org/10.48560/rspo.22582

Keywords:

Cornea, Corneal Diseases, Corneal Topography

Abstract

PURPOSE: To evaluate the inter-device agreement between Placido -based topography (Topolyzer Vario, WaveLight, Alcon Laboratories, Inc., Ft Worth, Texas, USA) and rotational Scheimpflug tomography (Oculyzer II, WaveLight, Alcon Laboratories, Inc., Ft Worth, Texas, USA) for measuring anterior corneal surface parameters.

METHODS: Fifty eyes from 50 subjects with no ocular disease were included in this study. The main outcome measurements were keratometric readings and anterior surface irregularity metrics derived from Fourier analysis (decentration, decentration axis and irregularity). Inter-device correlation and agreement between the two devices were assessed.

RESULTS:No statistically significant differences were detected among the parameters obtained by both devices (paired t test; p > 0.05). The 95% LoA between instruments were within the clinically relevant margins of discrepancy for the keratometric parameters. Seven (14%) and five (10%) eyes presented a difference greater than 0.5 D for the flat meridian (K1) and steep meridian (K2) parameters, respectively. Regarding corneal astigmatism, only three eyes (6%) presented a difference greater than 0.5 D between both devices. All the parameters selected from the Fourier analysis showed a lower level of correlation and a more extensive range of LoA compared to the keratometric variables.

CONCLUSION: Placido topography and Scheimpflug tomography showed good correlation and agreement for keratometric measurements, including corneal astigmatism. Based on the data derived from the Fourier analysis, the results recommend caution when selecting the data to plan the topography-guided customized ablation, especially in primary eyes, which calls for prospective studies, possibly using simulated models.

Downloads

Download data is not yet available.

References

Kanellopoulos AJ, Asimellis G. Keratoconus management: long-term stability of topography-guided normalization combined with high-fluence CXL stabilization (the Athens Protocol). J Refract Surg. 2014;30:88-93. doi: 10.3928/1081597X-20140120-03.

Kanellopoulos AJ, Binder PS. Management of corneal ectasia after LASIK with combined, same-day, topography-guided partial transepithelial PRK and collagen cross-linking: the athens protocol. J Refract Surg. 2011;27:323-31. doi: 10.3928/1081597X-20101105-01.

Koller T, Iseli HP, Donitzky C, Ing D, Papadopoulos N, Seiler T. Topography-guided surface ablation for forme fruste keratoconus. Ophthalmology. 2006;113:2198-202. doi: 10.1016/j.ophtha.2006.06.032.

Stojanovic A, Zhang J, Chen X, Nitter TA, Chen S, Wang Q. Topography-guided transepithelial surface ablation followed by corneal collagen cross-linking performed in a single combined procedure for the treatment of keratoconus and pellucid marginal degeneration. J Refract Surg. 2010;26:145-52. doi: 10.3928/1081597X-20100121-10.

Holland S, Lin DT, Tan JC. Topography-guided laser refractive surgery. Curr Opin Ophthalmol. 2013;24:302-9. doi: 10.1097/ICU.0b013e3283622a59.

Stulting RD, Fant BS, Group TCS, et al. Results of topography-guided laser in situ keratomileusis custom ablation treatment with a refractive excimer laser. J Cataract Refract Surg. 2016;42:11-8. doi: 10.1016/j.jcrs.2015.08.016.

Faria-Correia F, Ribeiro S, Monteiro T, Lopes BT, Salomao MQ, Ambrosio R, Jr. Topography-Guided Custom Photorefractive Keratectomy for Myopia in Primary Eyes With the WaveLight EX500 Platform. J Refract Surg. 2018;34:541-6. doi: 10.3928/1081597X-20180705-03.

Cummings AB, Mascharka N. Outcomes after topography-based LASIK and LASEK with the wavelight oculyzer and topolyzer platforms. J Refract Surg. 2010;26:478-85. doi: 10.3928/1081597X-20090814-05.

Ambrosio R, Jr., Belin MW. Imaging of the cornea: topography vs tomography. J Refract Surg. 2010;26:847-9. doi: 10.3928/1081597X-20101006-01.

Ambrosio R, Jr., Valbon BF, Faria-Correia F, Ramos I, Luz A. Scheimpflug imaging for laser refractive surgery. Curr Opin Ophthalmol. 2013;24:310-20. doi: 10.1097/ICU.0b013e3283622a94.

Salomao MQ, Esposito A, Dupps WJ, Jr. Advances in anterior segment imaging and analysis. Curr Opin Ophthalmol. 2009;20:324-32.

Cummings A, Durrie D, Gordon M, Williams R, Gow JA, Maus M. Prospective Evaluation of Outcomes in Patients Undergoing Treatment for Myopia Using the WaveLight Refractive Suite. J Refract Surg. 2017;33:322-8. doi: 10.3928/1081597X-20160926-01.

Thibos LN, Wheeler W, Horner D. Power vectors: an application of Fourier analysis to the description and statistical analysis of refractive error. Optom Vis Sci. 1997;74:367-75.

Hjortdal JO, Erdmann L, Bek T. Fourier analysis of video-keratographic data. A tool for separation of spherical, regular astigmatic and irregular astigmatic corneal power components. Ophthalmic Physiol Opt. 1995;15:171-85.

Ambrosio R, Jr., Klyce SD, Wilson SE. Corneal topographic and pachymetric screening of keratorefractive patients. J Refract Surg. 2003;19:24-9.

Ambrosio R, Jr., Nogueira LP, Caldas DL, Fontes BM, Luz A, Cazal JO, et al. Evaluation of corneal shape and biomechanics before LASIK. Int Ophthalmol Clin. 2011;51:11-38. doi: 10.1097/IIO.0b013e31820f1d2d.

Ambrosio R, Jr., Randleman JB. Screening for ectasia risk: what are we screening for and how should we screen for it? J Refract Surg. 2013;29:230-2. doi: 10.3928/1081597X-20130318-01.

Gonen T, Cosar CB, Sener B, Keskinbora KH. Comparison of keratometric data obtained by automated keratometer, Dicon CT 200, Allegro Topolyzer, and Pentacam. J Refract Surg. 2012;28:557-61. doi: 10.3928/1081597X-20120723-04.

Wang Q, Savini G, Hoffer KJ, Xu Z, Feng Y, Wen D, et al. A comprehensive assessment of the precision and agreement of anterior corneal power measurements obtained using 8 different devices. PLoS One. 2012;7:e45607. doi: 10.1371/journal.pone.0045607.

Kanellopoulos AJ, Asimellis G. Correlation between central corneal thickness, anterior chamber depth, and corneal keratometry as measured by Oculyzer II and WaveLight OB820 in preoperative cataract surgery patients. J Refract Surg. 2012;28:895-900. doi: 10.3928/1081597X-20121005-07.

Chen D, Lam AK. Reliability and repeatability of the Pentacam on corneal curvatures. Clin Exp Optom.. 2009;92:110-8.

Delrivo M, Ruiseñor Vázquez PR, Galletti JD, Garibotto M, Fuentes Bonthoux F, Pförtner T, et al. Agreement between placido topography and Scheimpflug tomography for corneal astigmatism assessment. J Refract Surg. 2014;30:49-53. doi: 10.3928/1081597x-20131217-06.

Savini G, Schiano-Lomoriello D, Hoffer KJ. Repeatability of automatic measurements by a new anterior segment optical coherence tomographer combined with Placido topography and agreement with 2 Scheimpflug cameras. J Cataract Refract Surg. 2018;44:471-8. doi: 10.1016/j.jcrs.2018.02.015.

Wang L, Chernyak D, Yeh D, Koch DD. Fitting behaviors of Fourier transform and Zernike polynomials. J Cataract Refract Surg. 2007;33:999-1004.

Yoon G, Pantanelli S, MacRae S. Comparison of Zernike and Fourier wavefront reconstruction algorithms in representing corneal aberration of normal and abnormal eyes. J Refract Surg 2008;24:582-90.

Liu Z, Pflugfelder SC. Corneal surface regularity and the effect of artificial tears in aqueous tear deficiency. Ophthalmology. 1999;106:939-43.

Rosales P, De Castro A, Jimenez-Alfaro I, Marcos S. Intraocular lens alignment from purkinje and Scheimpflug imaging. Clin Exp Optom. 2010;93:400-8. doi: 10.1111/j.1444-0938.2010.00514.x.

Seven I, Lloyd JS, Dupps WJ. Differences in Simulated Refractive Outcomes of Photorefractive Keratectomy (PRK) and Laser In-Situ Keratomileusis (LASIK) for Myopia in Same-Eye Virtual Trials. Int J Environ Res Public Health. 2020;17:287. doi: 10.3390/ijerph17010287.

Downloads

Published

2021-03-31

How to Cite

Faria-Correia, F., Ribeiro, S., Lopes, B. T., Salomão, M. Q., & Ambrósio Jr., R. (2021). Anterior Surface Measurements Correlation and Agreement Between Rotational Scheimpflug and Placido Disc-based Devices in Primary Eyes. Revista Sociedade Portuguesa De Oftalmologia, 45(1), 29–35. https://doi.org/10.48560/rspo.22582

Issue

Section

Original Article